- 1. Determining oxygen content in the environment of industrial and mining enterprises, cabins, tunnels, variable pressure suction nitrogen machines, oxygen generators, , and fertilizer industrial es is amazingly effective in avoiding system explosions due to oxygen in the production process and preventing room interest accidents in hypoxic environments..
- 2. The oxygen content in the exhaust of the internal combustion engine and the oxygen content in the flue is detected to achieve reasonable combustion control and save fuel..
- 3.Oxygen content determination is carried out in the atmosphere storage of grains, fruits, vegetables, , Chinese al medicines, tea leaves, dried fruits, etc..
- 4. Analysis of oxygen content in ventilators, anesthesia machines, hyperbaric oxygen chambers, infant incubators, and aerospace and diving operations in the medical business..
- 5. It is essential experimental equipment for physical chemistry, biology, and other experiments in secondary schools, colleges and universities, and medical schools. 6. It is suitable for determining oxygen content in the meteorological department electrolysis hydrogen production, air separation oxygen production, nitrogen production, and ceutical factory injection quality inspection..
The instrument's sensor adopts advanced oxygen electrode
manufacturing process technology, which achieves long life, high
stability, high precision, fast response time, and can resist the
interference of various harmful es (especially the
interference of hydrogen). The oxygen electrode produced by our
factory has also obtained Chinese patents. The instrument host is
assembled by an imported A/D conversion circuit and digital
display, which has the characteristics of easy operation and wide
range of 0.00%-100%, particularly suitable for the following
occasions.
The sensor of the instrument is an electro polarographic
diaphragm oxygen electrode. It uses a platinum cathode and a
silver-silver chloride anode. Potassium chloride is used as an
electrolyte, and the diaphragm material is a
polyperfluoroethylene propylene film selected to transmit oxygen.
The electrode reaction is as follows
Cathode: O ₂ + 2H ₂ O + 4e → 4OHˉ
Ag + Clˉ → AgCl
The electrode current is proportional to the oxygen content in
the , driven by analog conversion and decoding, and displays
the percentage of oxygen through a digital screen.
Technical performance
1. Range Accuracy
0-100% Minimum resolution is 1% ±3%.
0-50.0% The minimum resolution is 0.1% ±1%.
0-5.00% Minimum resolution is 0.01% ±0.5%.
2. Ambient temperature: 0-40°C, humidity <90%
3. Response time: 5 seconds to indicate 90% of the oxygen
concentration (20°C)
4. Sampling method: hand pinch suction type
5. Power supply: 9V lithium battery, can work continuously for
200 hours
6. Dimensions: 125×70×25 mm
7. Weight: 200g
8. Stability: 8 hours ±1%.
How to use Rotate the switch to 50 gear and stabilize for about 3
minutes (the number of just started up drops from large to small,
and the descent is the balancing process of the electrode, which
is a normal phenomenon). Pinch the suction balloon 2-3 times,
inhale the fresh air, and adjust the 21% twist so that the LCD
screen displays 21.0
After the above work is completed, the instrument can be on-site
detection, the sampling device is connected to the outlet to
be measured, the other end is connected to the suction balloon,
and the suction balloon is pinched 3-4 times. The oxygen content
in the measured can be obtained after the reading is stable.
If the measured value is higher than 50%, the switch is rotated
to 100%, and when it is lower than 5%, the button is turned to 5%
to improve data accuracy. After the inspection is complete,
rotate the switch to the OFF position.
The maintenance of the instrument
This instrument uses a diaphragm oxygen electrode in a clean
and relative humidity of about 85 environments, generally can be
used for two years left and right, such as electrode aging can
not be used, immediately send the instrument back to the factory
for repair, maintenance costs are calculated separately.
During use, the instrument's sensitivity decreases significantly
as the use time is extended, caused by matter and dust
contamination in the to be measured. This can be solved by
treating the cleanly during sampling, which is the key to
extending the life of the electrode.
Fault:
① If the instrument's response is slow during the test and there
is a significant drift after calibration, the principle is that
there is water vapor or water vapor on the electrode head.
Dust, please gently dab the water ball and dust with cotton wool
to remove.
② If the calibration cannot be performed: 1. The air
permeability is decreased due to the contamination of the
diaphragm. 2. There are sharp objects to scratch the diaphragm,
and the electrodes are aging.
③ If the LOBAT indicator in the upper right corner is on, the
battery voltage is insufficient, and the back cover should be
opened to replace the new battery.
Note: If the instrument is not used for a long time, please
remove the battery.